KEYCOM provides solutions in the 300 GHz to 1 μHz frequency range (SHF/EHF/MW/UHF/VHF/ULF).
We undertake designing, manufacturing and measuring. Please feel free to contact us.

S-parameter method Free space type
transmission mode for flat plate
dielectric constant and permeability
measurement system

Catalog No.dps21002040614-08

Unlike a coaxial tube type or a wave guide tube type, this method does not make errors caused by air-gap as specimens are not put into a fixture.
Practical data can be obtained even at the rough, uneven surfaces where the specimen is placed on.
It enables compactness and measuring specimens by plane wave because of direct installment of lens with an antenna.
It enables measurement by monitoring S21 and S11 parameter with connections of a test fixture to a vector network analyzer and PC.

Specifications

Frequency range 2.6-26.5GHz
Dielectric constant (εr') 2-30 Accuracy ±10%
Dielectric loss tangent (tanδ) 0.01-20
Permeability (μr') 1-100 Accuracy ±10%
Permeability (μr'') 0.01-20

Specimen size more than 450mm × 450mm
Specimen thickness Effective wavelength 1/4
Lens size 430mmφ

Model No.

Name Specifications Model No.
Main body   LA-430
Antenna (40 GHz or less in case of a coaxial wave guide tube conversion installed) Frequency 18-26.5GHz (WR-42) RH42S23SMA(f)5
Frequency 12.4-18GHz (WR-62) RH62S22SMA(f)5
Frequency 8.2-12.4GHz (WR-90) RH90S19SMA(f)5
Frequency 5.85-8.2GHz (WR-137) RH137S18SMA(f)5
Frequency 4.90-7.05GHz (WR-159) RH159S17SMA(f)5
Frequency 3.95-5.85GHz (WR-187) RH187S16SMA(f)5
Frequency 2.60-3.95GHz (WR-284) Rh284S14SMA(f)5
Software for calculation (for Windows PC) εr', tanδ, μr ', and μr" is displayed, on horizontal axis as a frequency.
DMP-15
Coaxial cable for mIcro wave, 2set (3m、40GHz type)    
GPIB cable    

Configurations

Main Body (Lens, Lens fixed frame, Antennna fixed stand, adjustment mechanism of specimen slight movement)